We report the status of the art of a prototype based on a GaAs pixel detector bump-bonded to a dedicated VLSI chip to be possibly used for imaging in the nuclear medicine field. This device, with a thick pixel matrix (64×64 square pixels, side), has already been tested with very good results for digital mammography applications (mean energy 20 keV). For more energetic photons, as in nuclear medicine, a thick detector has been chosen. Using radioactive sources (241Am, 60 keV and Tc, 140 keV photons) we have measured the performance of our prototype in terms of charge collection and detection efficiency of the detector, discrimination capability of the electronics and imaging properties of the whole system. In particular, we have evaluated the spatial resolution properties measuring the Point Spread Function and the imaging capabilities using a home made thyroid phantom. We present also the comparison between these results and those obtained with a traditional gamma camera and the evaluation, made by both experimental measurements and software simulations, of the geometry related to the use of a collimator.
Amendolia, S.R., Bisogni, M.G., Bottigli, U., Ceccopieri, A., Delogu, P., Dipasquale, G., et al. (2001). Performance of a medical imaging system for photons in the 60–140 keV energy range. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 461(1-3), 422-424 [10.1016/S0168-9002(00)01264-X].
Performance of a medical imaging system for photons in the 60–140 keV energy range
BOTTIGLI, U.;DELOGU, P.;MAESTRO, P.;
2001-01-01
Abstract
We report the status of the art of a prototype based on a GaAs pixel detector bump-bonded to a dedicated VLSI chip to be possibly used for imaging in the nuclear medicine field. This device, with a thick pixel matrix (64×64 square pixels, side), has already been tested with very good results for digital mammography applications (mean energy 20 keV). For more energetic photons, as in nuclear medicine, a thick detector has been chosen. Using radioactive sources (241Am, 60 keV and Tc, 140 keV photons) we have measured the performance of our prototype in terms of charge collection and detection efficiency of the detector, discrimination capability of the electronics and imaging properties of the whole system. In particular, we have evaluated the spatial resolution properties measuring the Point Spread Function and the imaging capabilities using a home made thyroid phantom. We present also the comparison between these results and those obtained with a traditional gamma camera and the evaluation, made by both experimental measurements and software simulations, of the geometry related to the use of a collimator.File | Dimensione | Formato | |
---|---|---|---|
NIMA461_tris.pdf
non disponibili
Tipologia:
Post-print
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
100.54 kB
Formato
Adobe PDF
|
100.54 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11365/27643
Attenzione
Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo