Background: Krebs von den Lungen-6 (KL-6) has been proposed as a disease severity marker of COVID-19. All research articles reported the KL-6 assay detected through Fujirebio reagents by Lumipulse G600/G1200 instrument. In the present study, KL-6 assay was analysed through Tosoh AIA-360 and compared with analytical results by Lumipulse G600 in a population of COVID-19 patients. Materials and methods: Sixty-four patients (median age, IQR 67 (58–76) years), all hos-pitalized for COVID-19 interstitial pneumonia at Siena COVID Unit. KL-6 was measured by two methods, chemiluminescence enzyme immunoassay (CLEIA) and fluorescent enzyme immunoassay (FEIA) method by Lumipulse G600 II and AIA 360 systems, respectively. Results: KL-6 concentrations evaluated by Lumipulse G600II were significantly higher in severe than those in non-severe patients (p < 0.0001) as well as evaluating by AIA360 (p < 0.0001). Receiver operating curve (ROC) curve analysis showed that KL-6 concentrations, by Lumipuse G600II, distinguished severe from non-severe COVID-19 patients with an area under the curve (AUC) of 99.8% and the best cut-off value was 448 U/mL. AUROC between severe and non-severe COVID-19 patients using T0 KL-6 concentrations by AIA360 was 97.4% and the best cut-off value was 398 U/mL. According to T0 KL-6 concentrations in COVID-19 patients, Bland–Altman difference analysis revealed a mean bias of 78 ± 174.8; while using T1 KL-6 concentrations in COVID-19 patients, Bland–Altman difference analysis revealed a mean bias of 48 ± 126 (95% limits of agreement −199–295) between the Lumipulse G600 II and the AIA360 systems. Conclusion: In conclusion, our study demonstrated that CLEIA and FEIA methods for serum KL-6 detection are comparable and reliable. KL-6 was confirmed as an easily detectable and effective biomarker to identify severe COVID-19 patients.

D'Alessandro, M., Bergantini, L., Cavallaro, D., Gangi, S., Cameli, P., Conticini, E., et al. (2022). Krebs von den Lungen-6 as Disease Severity Marker for COVID-19 Patients: Analytical Verification and Quality Assessment of the Tosoh AIA-360 Compared to Lumipulse G600II. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 19(4) [10.3390/ijerph19042176].

Krebs von den Lungen-6 as Disease Severity Marker for COVID-19 Patients: Analytical Verification and Quality Assessment of the Tosoh AIA-360 Compared to Lumipulse G600II

D'alessandro M.;Bergantini L.;Gangi S.;Cameli P.;Conticini E.;Frediani B.;Dotta F.;Bargagli E.;
2022-01-01

Abstract

Background: Krebs von den Lungen-6 (KL-6) has been proposed as a disease severity marker of COVID-19. All research articles reported the KL-6 assay detected through Fujirebio reagents by Lumipulse G600/G1200 instrument. In the present study, KL-6 assay was analysed through Tosoh AIA-360 and compared with analytical results by Lumipulse G600 in a population of COVID-19 patients. Materials and methods: Sixty-four patients (median age, IQR 67 (58–76) years), all hos-pitalized for COVID-19 interstitial pneumonia at Siena COVID Unit. KL-6 was measured by two methods, chemiluminescence enzyme immunoassay (CLEIA) and fluorescent enzyme immunoassay (FEIA) method by Lumipulse G600 II and AIA 360 systems, respectively. Results: KL-6 concentrations evaluated by Lumipulse G600II were significantly higher in severe than those in non-severe patients (p < 0.0001) as well as evaluating by AIA360 (p < 0.0001). Receiver operating curve (ROC) curve analysis showed that KL-6 concentrations, by Lumipuse G600II, distinguished severe from non-severe COVID-19 patients with an area under the curve (AUC) of 99.8% and the best cut-off value was 448 U/mL. AUROC between severe and non-severe COVID-19 patients using T0 KL-6 concentrations by AIA360 was 97.4% and the best cut-off value was 398 U/mL. According to T0 KL-6 concentrations in COVID-19 patients, Bland–Altman difference analysis revealed a mean bias of 78 ± 174.8; while using T1 KL-6 concentrations in COVID-19 patients, Bland–Altman difference analysis revealed a mean bias of 48 ± 126 (95% limits of agreement −199–295) between the Lumipulse G600 II and the AIA360 systems. Conclusion: In conclusion, our study demonstrated that CLEIA and FEIA methods for serum KL-6 detection are comparable and reliable. KL-6 was confirmed as an easily detectable and effective biomarker to identify severe COVID-19 patients.
2022
D'Alessandro, M., Bergantini, L., Cavallaro, D., Gangi, S., Cameli, P., Conticini, E., et al. (2022). Krebs von den Lungen-6 as Disease Severity Marker for COVID-19 Patients: Analytical Verification and Quality Assessment of the Tosoh AIA-360 Compared to Lumipulse G600II. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 19(4) [10.3390/ijerph19042176].
File in questo prodotto:
File Dimensione Formato  
Krebs.pdf

accesso aperto

Descrizione: Articolo
Tipologia: PDF editoriale
Licenza: Creative commons
Dimensione 1.71 MB
Formato Adobe PDF
1.71 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/1203645