PurposeTo evaluate and compare the diagnostic accuracy of the Humphrey Field Analyzer (HFA), Octopus perimetry, and Cirrus OCT for glaucomatous optic neuropathy.MethodsEighty-eight healthy individuals and 150 open-angle glaucoma patients were consecutive and prospectively selected. Eligibility criteria for the glaucoma group were intraocular pressure ≥21 mm Hg and glaucomatous optic nerve head morphology. All subjects underwent a reliable standard automated perimetry with the HFA and Octopus perimeter, and were imaged with the Cirrus OCT. Receiver-operating characteristic (ROC) curves were plotted for the threshold values and main indices of the HFA and Octopus, the peripapillary retinal nerve fiber layer thicknesses, and the optic nerve head parameters. Sensitivities at 85 and 95% fixed-specificities were also calculated. The best areas under the ROC curves (AUCs) were compared using the DeLong method.ResultsIn the glaucoma group, mean deviation (MD) was -5.42±4.6 dB for HFA and 3.90±3.6 dB for Octopus. The MD of the HFA (0.966; P<0.001), mean sensitivity of the Octopus (0.941; P<0.001), and average cup-to-disc (C/D) ratio measured by the Cirrus OCT (0.958; P<0.001) had the largest AUCs for each test studied. There were no significant differences among them. Sensitivities at 95% fixed-specificity were 82% for pattern standard deviation of the HFA, 81.3% for average C/D ratio of OCT, and 80% for the MD of the Octopus.ConclusionsHFA, Octopus, and Cirrus OCT demonstrated similar diagnostic accuracies for glaucomatous optic neuropathy. Visual field and OCT provide supplementary information and thus these tests are not interchangeable. © 2017 Macmillan Publishers Limited, part of Springer Nature. All rights reserved.

Monsalve, B., Ferreras, A., Calvo, P., Urcola, J.A., Figus, M., Monsalve, J., et al. (2017). Diagnostic ability of Humphrey perimetry, Octopus perimetry, and optical coherence tomography for glaucomatous optic neuropathy. EYE, 31(3), 443-451 [10.1038/eye.2016.251].

Diagnostic ability of Humphrey perimetry, Octopus perimetry, and optical coherence tomography for glaucomatous optic neuropathy

FREZZOTTI, PAOLO
2017-01-01

Abstract

PurposeTo evaluate and compare the diagnostic accuracy of the Humphrey Field Analyzer (HFA), Octopus perimetry, and Cirrus OCT for glaucomatous optic neuropathy.MethodsEighty-eight healthy individuals and 150 open-angle glaucoma patients were consecutive and prospectively selected. Eligibility criteria for the glaucoma group were intraocular pressure ≥21 mm Hg and glaucomatous optic nerve head morphology. All subjects underwent a reliable standard automated perimetry with the HFA and Octopus perimeter, and were imaged with the Cirrus OCT. Receiver-operating characteristic (ROC) curves were plotted for the threshold values and main indices of the HFA and Octopus, the peripapillary retinal nerve fiber layer thicknesses, and the optic nerve head parameters. Sensitivities at 85 and 95% fixed-specificities were also calculated. The best areas under the ROC curves (AUCs) were compared using the DeLong method.ResultsIn the glaucoma group, mean deviation (MD) was -5.42±4.6 dB for HFA and 3.90±3.6 dB for Octopus. The MD of the HFA (0.966; P<0.001), mean sensitivity of the Octopus (0.941; P<0.001), and average cup-to-disc (C/D) ratio measured by the Cirrus OCT (0.958; P<0.001) had the largest AUCs for each test studied. There were no significant differences among them. Sensitivities at 95% fixed-specificity were 82% for pattern standard deviation of the HFA, 81.3% for average C/D ratio of OCT, and 80% for the MD of the Octopus.ConclusionsHFA, Octopus, and Cirrus OCT demonstrated similar diagnostic accuracies for glaucomatous optic neuropathy. Visual field and OCT provide supplementary information and thus these tests are not interchangeable. © 2017 Macmillan Publishers Limited, part of Springer Nature. All rights reserved.
2017
EYE
Monsalve, B., Ferreras, A., Calvo, P., Urcola, J.A., Figus, M., Monsalve, J., et al. (2017). Diagnostic ability of Humphrey perimetry, Octopus perimetry, and optical coherence tomography for glaucomatous optic neuropathy. EYE, 31(3), 443-451 [10.1038/eye.2016.251].
File in questo prodotto:
File Dimensione Formato  
Diagnostic ability of Humphrey perimetry-Monsalve-2017.pdf

non disponibili

Tipologia: PDF editoriale
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 1.84 MB
Formato Adobe PDF
1.84 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/999486