PURPOSE. To evaluate the repeatability and agreement of 2 Scheimpflug tomographies (Pentacam HR, Oculus GmbH; Sirius, CSO Inc.) in measuring the central corneal thickness (CCT) and depth (ACD), volume (ACV), and angle (ACA) of the anterior chamber METHODS. Three consecutive measurements were performed in one eye of each subject, with both instruments. The following were evaluated: CCT, ACD, ACV, and ACA. Repeatability was assessed by the Friedman test and agreement was assessed by 4 different Wilcoxon tests for paired data. RESULTS. The study enrolled 30 patients. With Sirius tomography the Friedman test p values for testing homogeneity across measurements was with respect to CCT 0.320, ACD 0.315, ACV 0.171, and ACA 0.020; and with Pentacam CCT 0.191, ACD 0.010, ACV 0.704, and ACA 0.150. The Wilcoxon test p values for testing the measurement homogeneity between the corneal tomography Sirius and Pentacam showed the following: ACD 0.000, ACD 0.000, ACV 0.853, and ACA 0.000. CONCLUSIONS. The measurements may be considered substantially stable for each corneal tomography and each variable, but the 2 corneal tomographies do not produce statistically equivalent measurements.

Bedei, A., Appolloni, I., Madesani, A., Pietrelli, A., Franceschi, S., Barabesi, L. (2012). Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 22(Supplemento 7), 29-32 [10.5301/ejo.5000102].

Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth

Franceschi, Sara;Barabesi, Lucio
2012-01-01

Abstract

PURPOSE. To evaluate the repeatability and agreement of 2 Scheimpflug tomographies (Pentacam HR, Oculus GmbH; Sirius, CSO Inc.) in measuring the central corneal thickness (CCT) and depth (ACD), volume (ACV), and angle (ACA) of the anterior chamber METHODS. Three consecutive measurements were performed in one eye of each subject, with both instruments. The following were evaluated: CCT, ACD, ACV, and ACA. Repeatability was assessed by the Friedman test and agreement was assessed by 4 different Wilcoxon tests for paired data. RESULTS. The study enrolled 30 patients. With Sirius tomography the Friedman test p values for testing homogeneity across measurements was with respect to CCT 0.320, ACD 0.315, ACV 0.171, and ACA 0.020; and with Pentacam CCT 0.191, ACD 0.010, ACV 0.704, and ACA 0.150. The Wilcoxon test p values for testing the measurement homogeneity between the corneal tomography Sirius and Pentacam showed the following: ACD 0.000, ACD 0.000, ACV 0.853, and ACA 0.000. CONCLUSIONS. The measurements may be considered substantially stable for each corneal tomography and each variable, but the 2 corneal tomographies do not produce statistically equivalent measurements.
2012
Bedei, A., Appolloni, I., Madesani, A., Pietrelli, A., Franceschi, S., Barabesi, L. (2012). Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 22(Supplemento 7), 29-32 [10.5301/ejo.5000102].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/956443
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo