Echocardiography is a valuable technique to assess cardiac output (CO) in trauma patients, but it does not allow a continuous bedside monitoring. Beat-to-beat CO assessment can be obtained by other techniques, including the pulse contour method MostCare. The aim of our study was to compare CO obtained with MostCare (MC-CO) with CO estimated by transthoracic echocardiography (TTE-CO) in trauma patients. METHODS: Forty-nine patients with blunt trauma admitted to an intensive care unit and requiring hemodynamic optimization within 24 hours from admission were studied. TTE-CO and MC-CO were estimated simultaneously at baseline, after a fluid challenge and after the start of vasoactive drug therapy. RESULTS: One hundred sixteen paired CO values were obtained. TTE-CO values ranged from 2.9 to 7.6 L·min-1, and MC-CO ranged from 2.8 to 8.2 L·min-1. The correlation between the two methods was 0.94 (95% confidence interval [CI] = 0.89 to 0.97; p<0.001). The mean bias was -0.06 L·min-1 with limits of agreements (LoA) of -0.94 to 0.82 L·min-1 (lower 95% CI, -1.16 to -0.72; upper 95% CI, 0.60 to 1.04) and a percentage error of 18%. Changes in CO showed a correlation of 0.91 (95% CI = 0.87 to 0.95; p<0.001), a mean bias of - 0.01 L·min-1 with LoA of -0.67 to 0.65 L·min-1 (lower 95% CI, -0.83 to -0.51; upper 95% CI, 0.48 to 0.81). CONCLUSION: CO measured by MostCare showed good agreement with CO obtained by transthoracic echocardiography. Pulse contour analysis can complement echocardiography in evaluating hemodynamics in trauma patients.

Franchi, F., Falciani, E., Donadello, K., Zacà, V., Silvestri, R., Taccone, F.s., et al. (2013). Echocardiography and pulse contour analysis to assess cardiac output in trauma patients. MINERVA ANESTESIOLOGICA, 79, 137-146.

Echocardiography and pulse contour analysis to assess cardiac output in trauma patients.

FRANCHI, FEDERICO;SCOLLETTA, SABINO
2013-01-01

Abstract

Echocardiography is a valuable technique to assess cardiac output (CO) in trauma patients, but it does not allow a continuous bedside monitoring. Beat-to-beat CO assessment can be obtained by other techniques, including the pulse contour method MostCare. The aim of our study was to compare CO obtained with MostCare (MC-CO) with CO estimated by transthoracic echocardiography (TTE-CO) in trauma patients. METHODS: Forty-nine patients with blunt trauma admitted to an intensive care unit and requiring hemodynamic optimization within 24 hours from admission were studied. TTE-CO and MC-CO were estimated simultaneously at baseline, after a fluid challenge and after the start of vasoactive drug therapy. RESULTS: One hundred sixteen paired CO values were obtained. TTE-CO values ranged from 2.9 to 7.6 L·min-1, and MC-CO ranged from 2.8 to 8.2 L·min-1. The correlation between the two methods was 0.94 (95% confidence interval [CI] = 0.89 to 0.97; p<0.001). The mean bias was -0.06 L·min-1 with limits of agreements (LoA) of -0.94 to 0.82 L·min-1 (lower 95% CI, -1.16 to -0.72; upper 95% CI, 0.60 to 1.04) and a percentage error of 18%. Changes in CO showed a correlation of 0.91 (95% CI = 0.87 to 0.95; p<0.001), a mean bias of - 0.01 L·min-1 with LoA of -0.67 to 0.65 L·min-1 (lower 95% CI, -0.83 to -0.51; upper 95% CI, 0.48 to 0.81). CONCLUSION: CO measured by MostCare showed good agreement with CO obtained by transthoracic echocardiography. Pulse contour analysis can complement echocardiography in evaluating hemodynamics in trauma patients.
2013
Franchi, F., Falciani, E., Donadello, K., Zacà, V., Silvestri, R., Taccone, F.s., et al. (2013). Echocardiography and pulse contour analysis to assess cardiac output in trauma patients. MINERVA ANESTESIOLOGICA, 79, 137-146.
File in questo prodotto:
File Dimensione Formato  
R02Y2013N02A0137.pdf

accesso aperto

Descrizione: Articolo Principale
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 365.9 kB
Formato Adobe PDF
365.9 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/41901
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo