This study investigated the influence of the light guide shape on the polymerization efficacy of a light-activated resin-based composite cured with LED units as a function of the distance between the tip and the restoration. Two different LED units, each with different light guides and shapes, were used. Their ability to cure a single restorative material was investigated. The efficacy of the light-curing system was evaluated by measuring the depth of cure using the Acetone Shake test. Considering the ratio (R) between the entry diameter and the exit diameter of the light guide, the tips with a higher R-value proved to be more efficient if the tip and composite distance (D) was less than 5 mm, while for D > 5 mm, the tips with a lower R-value showed better results. The tip geometry of the tested light guide had a significant influence on the depth of cure of the tested resin composites. Therefore, depending on the distance, the more suitable light guide should be selected, based on the clinical situation. An R-value may be a better descriptor for the light guide shape than words such as "normal" or "turbo."

Corciolani, G., Vichi, A., Davidson, C.l., Ferrari, M. (2008). The influence of tip geometry and distance on light-curing efficacy. OPERATIVE DENTISTRY, 33(3), 325-331.

The influence of tip geometry and distance on light-curing efficacy.

VICHI, ALESSANDRO;FERRARI, MARCO
2008-01-01

Abstract

This study investigated the influence of the light guide shape on the polymerization efficacy of a light-activated resin-based composite cured with LED units as a function of the distance between the tip and the restoration. Two different LED units, each with different light guides and shapes, were used. Their ability to cure a single restorative material was investigated. The efficacy of the light-curing system was evaluated by measuring the depth of cure using the Acetone Shake test. Considering the ratio (R) between the entry diameter and the exit diameter of the light guide, the tips with a higher R-value proved to be more efficient if the tip and composite distance (D) was less than 5 mm, while for D > 5 mm, the tips with a lower R-value showed better results. The tip geometry of the tested light guide had a significant influence on the depth of cure of the tested resin composites. Therefore, depending on the distance, the more suitable light guide should be selected, based on the clinical situation. An R-value may be a better descriptor for the light guide shape than words such as "normal" or "turbo."
2008
Corciolani, G., Vichi, A., Davidson, C.l., Ferrari, M. (2008). The influence of tip geometry and distance on light-curing efficacy. OPERATIVE DENTISTRY, 33(3), 325-331.
File in questo prodotto:
File Dimensione Formato  
2008 Oper Dent. The influence of tip geometry and distance on light-curing efficacy.pdf

non disponibili

Tipologia: Post-print
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 725.19 kB
Formato Adobe PDF
725.19 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/32767
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo