The Abbott Real-Time HIV-1 assay was evaluated for its performance in quantification of human immunodeficiency virus type 1 (HIV-1) RNA in dried blood spot (DBS) samples. In total, 169 blood samples with detectable plasma HIV-1 RNA were used to extract RNA from paired DBS and liquid plasma samples, using the automated Abbott m Sample Preparation System (m2000sp). HIV-1 RNA was then quantitated by the m2000rt RealTime analyser. RNA samples suitable for real-time PCR were obtained from all but one (99.4%) of the DBS samples and HIV-1 RNA was detected in 163/168 (97.0%) samples. The correlation between HIV-1 RNA values measured in paired DBS and plasma samples was very high (r = 0.882), with 78.5% and 99.4% of cases differing by <0.5 and 1.0 log, respectively. Retesting of DBS replicates following 6 months of storage at 2-8 degrees C showed no loss of HIV-1 RNA in a subset of 89 samples. The feasibility of DBS testing coupled with automated sample processing, and the use of a latest-generation FDA-approved real-time PCR-based system, represents an encouraging first step for viral load measurement in reference centres in developing countries where access to antiretroviral therapy is expanding.

Marconi, A., Balestrieri, M., Comastri, G., Pulvirenti, F.R., Gennari, W., Tagliazucchi, S., et al. (2009). Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens. CLINICAL MICROBIOLOGY AND INFECTION, 15(1), 93-97 [10.1111/j.1469-0691.2008.02116.x].

Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens

ZAZZI M.
2009-01-01

Abstract

The Abbott Real-Time HIV-1 assay was evaluated for its performance in quantification of human immunodeficiency virus type 1 (HIV-1) RNA in dried blood spot (DBS) samples. In total, 169 blood samples with detectable plasma HIV-1 RNA were used to extract RNA from paired DBS and liquid plasma samples, using the automated Abbott m Sample Preparation System (m2000sp). HIV-1 RNA was then quantitated by the m2000rt RealTime analyser. RNA samples suitable for real-time PCR were obtained from all but one (99.4%) of the DBS samples and HIV-1 RNA was detected in 163/168 (97.0%) samples. The correlation between HIV-1 RNA values measured in paired DBS and plasma samples was very high (r = 0.882), with 78.5% and 99.4% of cases differing by <0.5 and 1.0 log, respectively. Retesting of DBS replicates following 6 months of storage at 2-8 degrees C showed no loss of HIV-1 RNA in a subset of 89 samples. The feasibility of DBS testing coupled with automated sample processing, and the use of a latest-generation FDA-approved real-time PCR-based system, represents an encouraging first step for viral load measurement in reference centres in developing countries where access to antiretroviral therapy is expanding.
2009
Marconi, A., Balestrieri, M., Comastri, G., Pulvirenti, F.R., Gennari, W., Tagliazucchi, S., et al. (2009). Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens. CLINICAL MICROBIOLOGY AND INFECTION, 15(1), 93-97 [10.1111/j.1469-0691.2008.02116.x].
File in questo prodotto:
File Dimensione Formato  
2009CMI_a.PDF

non disponibili

Tipologia: Post-print
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 77.7 kB
Formato Adobe PDF
77.7 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/25965
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo