The detachment of components of endovascular devices is a rare but potentially harmful complication. This report describes a clinical case of nosecone detachment from an iliac branch device delivery system during an aortoiliac aneurysm exclusion in a 73-year-old male patient. After correct deployment, the iliac branch device nosecone detached during the removal of the delivery system. The guidewire was snared from the contralateral side, and two long sheaths were advanced to trap and protect the fractured piece, ensuring nice alignment and enabling its safe removal. This case highlights key strategies for managing serious complications encountered during endovascular aortic repair. © 2025 The Author(s)

Pasquetti, L., Pasqui, E., Galzerano, G., Gargiulo, B., Molino, C., De Donato, G. (2025). Protected retrieval of a detached iliac branch device nosecone. JOURNAL OF VASCULAR SURGERY CASES AND INNOVATIVE TECHNIQUES, 11(4) [10.1016/j.jvscit.2025.101835].

Protected retrieval of a detached iliac branch device nosecone

Pasquetti L.
;
Pasqui E.;Galzerano G.;Gargiulo B.;Molino C.;de Donato G.
2025-01-01

Abstract

The detachment of components of endovascular devices is a rare but potentially harmful complication. This report describes a clinical case of nosecone detachment from an iliac branch device delivery system during an aortoiliac aneurysm exclusion in a 73-year-old male patient. After correct deployment, the iliac branch device nosecone detached during the removal of the delivery system. The guidewire was snared from the contralateral side, and two long sheaths were advanced to trap and protect the fractured piece, ensuring nice alignment and enabling its safe removal. This case highlights key strategies for managing serious complications encountered during endovascular aortic repair. © 2025 The Author(s)
2025
Pasquetti, L., Pasqui, E., Galzerano, G., Gargiulo, B., Molino, C., De Donato, G. (2025). Protected retrieval of a detached iliac branch device nosecone. JOURNAL OF VASCULAR SURGERY CASES AND INNOVATIVE TECHNIQUES, 11(4) [10.1016/j.jvscit.2025.101835].
File in questo prodotto:
File Dimensione Formato  
Protected retrieval of a detached iliac branch device nosecone.pdf

accesso aperto

Descrizione: Articolo
Tipologia: PDF editoriale
Licenza: Creative commons
Dimensione 1.15 MB
Formato Adobe PDF
1.15 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/1294095