Background: Neurofilament light chains (NfL) represent reliable serum biomarkers of neuroaxonal injury. Due to their low serum levels, precise detection methods are critical. This study aimed to scrutinize the comparability of two techniques: Single Molecule Array (SiMoA) and Ella automated immunoassay, analyzing serum NfL levels in ATTRv presymptomatic subjects and polyneuropathy patients. Methods: Blood samples were processed and analyzed using commercial Ella and SiMoA kits. Statistical analysis included the Wilcoxon signed-rank test, Spearman correlation, Bland-Altman, and Passing-Bablok regression. ANCOVA models were used to compare NfL levels between cohorts. Results: The study measured serum NfL levels in 55 symptomatic and 55 presymptomatic ATTRv subjects. Median NfL concentrations were significantly higher with Ella (median 27.5 pg/mL) than SiMoA (median 15.9 pg/mL). Both methods showed a strong positive correlation (r = 0.8, p < 0.001), but Ella overestimated NfL by 42%. Bland-Altman analysis revealed a mean bias of 15.4 pg/mL, with limits of agreement between -41.1 and 72.0 pg/mL. The slope of the Passing-Bablok regression line was 0.58, and the intercept was 3.48 pg/mL, suggesting that the Ella platform tends to overestimate NfL concentrations compared to the SiMoA platform, especially at higher concentrations. Both methods effectively distinguished presymptomatic from symptomatic patients (p < 0.001 for both). Conclusions: Our findings underscore that both platforms are effective in measuring serum NfL, with Ella consistently overestimating, especially at higher concentrations. The difference between the two platforms must be taken into account when deeming the concentrations as pathological or normal, as well as when conducting longitudinal studies.

Primiano, G., Luigetti, M., Righi, D., Romano, A., Leonardi, L., Guglielmino, V., et al. (2025). Comparative Analysis of SiMoA and Ella Immunoassay Platforms for Measuring Serum Neurofilament Light Chain Levels in ATTRv With Polyneuropathy and Presymptomatic Carriers. EUROPEAN JOURNAL OF NEUROLOGY, 32(6) [10.1111/ene.70215].

Comparative Analysis of SiMoA and Ella Immunoassay Platforms for Measuring Serum Neurofilament Light Chain Levels in ATTRv With Polyneuropathy and Presymptomatic Carriers

Righi, Delia;Manco, Carlo;Stromillo, Maria Laura;De Stefano, Nicola;Plantone, Domenico
2025-01-01

Abstract

Background: Neurofilament light chains (NfL) represent reliable serum biomarkers of neuroaxonal injury. Due to their low serum levels, precise detection methods are critical. This study aimed to scrutinize the comparability of two techniques: Single Molecule Array (SiMoA) and Ella automated immunoassay, analyzing serum NfL levels in ATTRv presymptomatic subjects and polyneuropathy patients. Methods: Blood samples were processed and analyzed using commercial Ella and SiMoA kits. Statistical analysis included the Wilcoxon signed-rank test, Spearman correlation, Bland-Altman, and Passing-Bablok regression. ANCOVA models were used to compare NfL levels between cohorts. Results: The study measured serum NfL levels in 55 symptomatic and 55 presymptomatic ATTRv subjects. Median NfL concentrations were significantly higher with Ella (median 27.5 pg/mL) than SiMoA (median 15.9 pg/mL). Both methods showed a strong positive correlation (r = 0.8, p < 0.001), but Ella overestimated NfL by 42%. Bland-Altman analysis revealed a mean bias of 15.4 pg/mL, with limits of agreement between -41.1 and 72.0 pg/mL. The slope of the Passing-Bablok regression line was 0.58, and the intercept was 3.48 pg/mL, suggesting that the Ella platform tends to overestimate NfL concentrations compared to the SiMoA platform, especially at higher concentrations. Both methods effectively distinguished presymptomatic from symptomatic patients (p < 0.001 for both). Conclusions: Our findings underscore that both platforms are effective in measuring serum NfL, with Ella consistently overestimating, especially at higher concentrations. The difference between the two platforms must be taken into account when deeming the concentrations as pathological or normal, as well as when conducting longitudinal studies.
2025
Primiano, G., Luigetti, M., Righi, D., Romano, A., Leonardi, L., Guglielmino, V., et al. (2025). Comparative Analysis of SiMoA and Ella Immunoassay Platforms for Measuring Serum Neurofilament Light Chain Levels in ATTRv With Polyneuropathy and Presymptomatic Carriers. EUROPEAN JOURNAL OF NEUROLOGY, 32(6) [10.1111/ene.70215].
File in questo prodotto:
File Dimensione Formato  
Euro J of Neurology - 2025 - Primiano - Comparative Analysis of SiMoA and Ella Immunoassay Platforms for Measuring Serum.pdf

accesso aperto

Descrizione: Articolo
Tipologia: PDF editoriale
Licenza: Creative commons
Dimensione 737.92 kB
Formato Adobe PDF
737.92 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/1293914