Objective: The objective of this study was to compare outcomes of patients with favorable neck angulation (FNA) and hostile neck angulation (HNA) treated with the Treovance stent graft (Terumo Aortic [formerly Bolton Medical], Sunrise, Fla). Methods: Patients with abdominal aortic aneurysms suitable for endovascular repair with Treovance were included in the RATIONALE postmarket surveillance registry. A post hoc subgroup analysis compared patients with infrarenal neck angles <60 degrees (FNA) and ≥60 degrees (HNA). Results: After 1 year, 179 FNA (89.5%) and 21 HNA (10.5%) patients were analyzed. Both groups were similar in terms of sex (male, 92.7% FNA and 95.2% HNA) and age (73.0 years vs 72.6 years), but the HNA group had more Asian or other race representation (7.3% vs 19.0%) and more patients assigned to American Society of Anesthesiologists class 3 and class 4 (57.6% vs 66.7%). Mean suprarenal angles (standard deviation) were 13.1 (±13.5) degrees vs 29.0 (±16.4) degrees; mean infrarenal angles were 23.2 (±16.4) degrees vs 65.4 (±4.6) degrees, respectively. Aneurysm sac size maximum diameter was 58.1 (±9.8) mm vs 62.0 (±14.1) mm. There was a significant difference in unplanned adjunctive procedures (2.2% vs 19.0%; P =.01). Mean procedural duration was also significantly different for HNA patients, who underwent protracted operations (111.3 [±47.3] minutes vs 153.5 [±44.5] minutes; P <.0001). However, there were no significant differences in rates of clinical success (96.1% vs 95.2%). The rate of reintervention was low overall but 0% in the HNA group. Changes in sac size at 1 year were significant in both groups but not as pronounced in HNA patients (relative change of −11.8% [±13.3] vs −6.6% [±11.4]). Conclusions: Patients with high neck angulation treated with Treovance underwent more complex procedures but showed equally good technical success and 1-year clinical success parameters.

Murray, D., Szeberin, Z., Benevento, D., Abdallah, F., Palasciano, G., Lescan, M., et al. (2020). A comparison of clinical outcomes of abdominal aortic aneurysm patients with favorable and hostile neck angulation treated by endovascular repair with the Treovance stent graft. JOURNAL OF VASCULAR SURGERY, 71(6), 1881-1889 [10.1016/j.jvs.2019.07.096].

A comparison of clinical outcomes of abdominal aortic aneurysm patients with favorable and hostile neck angulation treated by endovascular repair with the Treovance stent graft

Palasciano G.;Setacci C.
2020-01-01

Abstract

Objective: The objective of this study was to compare outcomes of patients with favorable neck angulation (FNA) and hostile neck angulation (HNA) treated with the Treovance stent graft (Terumo Aortic [formerly Bolton Medical], Sunrise, Fla). Methods: Patients with abdominal aortic aneurysms suitable for endovascular repair with Treovance were included in the RATIONALE postmarket surveillance registry. A post hoc subgroup analysis compared patients with infrarenal neck angles <60 degrees (FNA) and ≥60 degrees (HNA). Results: After 1 year, 179 FNA (89.5%) and 21 HNA (10.5%) patients were analyzed. Both groups were similar in terms of sex (male, 92.7% FNA and 95.2% HNA) and age (73.0 years vs 72.6 years), but the HNA group had more Asian or other race representation (7.3% vs 19.0%) and more patients assigned to American Society of Anesthesiologists class 3 and class 4 (57.6% vs 66.7%). Mean suprarenal angles (standard deviation) were 13.1 (±13.5) degrees vs 29.0 (±16.4) degrees; mean infrarenal angles were 23.2 (±16.4) degrees vs 65.4 (±4.6) degrees, respectively. Aneurysm sac size maximum diameter was 58.1 (±9.8) mm vs 62.0 (±14.1) mm. There was a significant difference in unplanned adjunctive procedures (2.2% vs 19.0%; P =.01). Mean procedural duration was also significantly different for HNA patients, who underwent protracted operations (111.3 [±47.3] minutes vs 153.5 [±44.5] minutes; P <.0001). However, there were no significant differences in rates of clinical success (96.1% vs 95.2%). The rate of reintervention was low overall but 0% in the HNA group. Changes in sac size at 1 year were significant in both groups but not as pronounced in HNA patients (relative change of −11.8% [±13.3] vs −6.6% [±11.4]). Conclusions: Patients with high neck angulation treated with Treovance underwent more complex procedures but showed equally good technical success and 1-year clinical success parameters.
2020
Murray, D., Szeberin, Z., Benevento, D., Abdallah, F., Palasciano, G., Lescan, M., et al. (2020). A comparison of clinical outcomes of abdominal aortic aneurysm patients with favorable and hostile neck angulation treated by endovascular repair with the Treovance stent graft. JOURNAL OF VASCULAR SURGERY, 71(6), 1881-1889 [10.1016/j.jvs.2019.07.096].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/1142943
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo