Purpose: To evaluate the safety and performance of the Treovance stent-graft. Methods: The global, multicenter RATIONALE registry (ClinicalTrials.gov; identifier NCT03449875) prospectively enrolled 202 patients (mean age 73.0±7.8 years; 187 men) with abdominal aortic aneurysms (AAA) suitable for endovascular aneurysm repair (EVAR) using the Treovance. The composite primary safety endpoint was site-reported all-cause mortality and major morbidity. The primary efficacy outcome was clinical success. Further outcomes evaluated included technical success; stent-graft migration, patency, and integrity; endoleak; and aneurysm size changes. Results: Technical success was 96% (194/202); 8 patients had unresolved type I endoleaks at the end of the procedure. There was no 30-day mortality and 1% major morbidity (1 myocardial infarction and 1 bowel ischemia). Clinical success at 1 year was confirmed in 194 (96%) patients; 6 of 8 patients had new/persistent endoleaks and 2 had aneurysm expansion without identified endoleak. A total of 8 (4%) reinterventions were required during the mean 13.7±3.1 months of follow-up (median 12.8). At 1 year, the Kaplan-Meier estimate for freedom from reintervention was 95.6% (95% CI 91.4% to 97.8%). Other estimates were 95.5% (95% CI 91.7% to 97.6%) for freedom from endoleak type I/III and 97.4% (95% CI 94.2% to 98.9%) for freedom from aneurysm expansion. Thirteen (6.4%) patients died; no death was aneurysm related. Conclusion: The RATIONALE registry showed favorable safety and clinical performance of the Treovance stent-graft for the treatment of infrarenal AAAs in a real-world setting.
Uberoi, R., Setacci, C., Lescan, M., Lorido, A., Murray, D., Szeberin, Z., et al. (2018). Global Post-Market Clinical Follow-up of the Treovance Stent-Graft for Endovascular Aneurysm Repair: One-Year Results From the RATIONALE Registry. JOURNAL OF ENDOVASCULAR THERAPY, 25(6), 726-734 [10.1177/1526602818803939].
Global Post-Market Clinical Follow-up of the Treovance Stent-Graft for Endovascular Aneurysm Repair: One-Year Results From the RATIONALE Registry
Setacci C.;Palasciano G.;
2018-01-01
Abstract
Purpose: To evaluate the safety and performance of the Treovance stent-graft. Methods: The global, multicenter RATIONALE registry (ClinicalTrials.gov; identifier NCT03449875) prospectively enrolled 202 patients (mean age 73.0±7.8 years; 187 men) with abdominal aortic aneurysms (AAA) suitable for endovascular aneurysm repair (EVAR) using the Treovance. The composite primary safety endpoint was site-reported all-cause mortality and major morbidity. The primary efficacy outcome was clinical success. Further outcomes evaluated included technical success; stent-graft migration, patency, and integrity; endoleak; and aneurysm size changes. Results: Technical success was 96% (194/202); 8 patients had unresolved type I endoleaks at the end of the procedure. There was no 30-day mortality and 1% major morbidity (1 myocardial infarction and 1 bowel ischemia). Clinical success at 1 year was confirmed in 194 (96%) patients; 6 of 8 patients had new/persistent endoleaks and 2 had aneurysm expansion without identified endoleak. A total of 8 (4%) reinterventions were required during the mean 13.7±3.1 months of follow-up (median 12.8). At 1 year, the Kaplan-Meier estimate for freedom from reintervention was 95.6% (95% CI 91.4% to 97.8%). Other estimates were 95.5% (95% CI 91.7% to 97.6%) for freedom from endoleak type I/III and 97.4% (95% CI 94.2% to 98.9%) for freedom from aneurysm expansion. Thirteen (6.4%) patients died; no death was aneurysm related. Conclusion: The RATIONALE registry showed favorable safety and clinical performance of the Treovance stent-graft for the treatment of infrarenal AAAs in a real-world setting.File | Dimensione | Formato | |
---|---|---|---|
Global Post-Market Clinical Follow-up of.pdf
accesso aperto
Tipologia:
PDF editoriale
Licenza:
Creative commons
Dimensione
338.68 kB
Formato
Adobe PDF
|
338.68 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11365/1134878
Attenzione
Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo