Introduction: During veno-venous extracorporeal membrane oxygenation, cardiac output monitoring is essential to assess tissue oxygen delivery. Adequate arterial oxygenation depends on the ratio between the extracorporeal pump blood flow and the cardiac output. The aim of this study was to compare estimates of cardiac output and blood flow/cardiac output ratios made using an uncalibrated pulse contour method with those made using echocardiography in patients treated with veno-venous extracorporeal membrane oxygenation. Methods: Cardiac output was estimated simultaneously using a pulse contour method (MostCareUp; Vygon, Encouen, France) and echocardiography in 17 hemodynamically stable patients treated with veno-venous extracorporeal membrane oxygenation. Comparisons were made using Bland–Altman and linear regression analysis. Results: There were significant correlations between cardiac output estimated using pulse contour method and echocardiography and between blood flow/cardiac output estimated using pulse contour method and blood flow/cardiac output estimated using echocardiography (r = 0.84, p < 0.001 and r = 0.87, p < 0.001, respectively). Bland–Altman analysis showed a good agreement (bias −0.20 ± 0.50 L/min) and a low percentage of error (25%) for the cardiac output values estimated by the two methods. The bias between the blood flow/cardiac output ratios obtained with the two methods was 5.19% ± 12.3% (percentage of error = 28.1%). Conclusions: The pulse contour method is a valuable alternative to echocardiography for the assessment of cardiac output and the blood flow/cardiac output ratio in patients treated with veno-venous extracorporeal membrane oxygenation.

Bond, O., Pozzebon, S., Franchi, F., Zama Cavicchi, F., Creteur, J., Vincent, J.-., et al. (2020). Comparison of estimation of cardiac output using an uncalibrated pulse contour method and echocardiography during veno-venous extracorporeal membrane oxygenation. PERFUSION-UK, 35(5), 397-401 [10.1177/0267659119883204].

Comparison of estimation of cardiac output using an uncalibrated pulse contour method and echocardiography during veno-venous extracorporeal membrane oxygenation

Franchi F.;Scolletta S.
2020-01-01

Abstract

Introduction: During veno-venous extracorporeal membrane oxygenation, cardiac output monitoring is essential to assess tissue oxygen delivery. Adequate arterial oxygenation depends on the ratio between the extracorporeal pump blood flow and the cardiac output. The aim of this study was to compare estimates of cardiac output and blood flow/cardiac output ratios made using an uncalibrated pulse contour method with those made using echocardiography in patients treated with veno-venous extracorporeal membrane oxygenation. Methods: Cardiac output was estimated simultaneously using a pulse contour method (MostCareUp; Vygon, Encouen, France) and echocardiography in 17 hemodynamically stable patients treated with veno-venous extracorporeal membrane oxygenation. Comparisons were made using Bland–Altman and linear regression analysis. Results: There were significant correlations between cardiac output estimated using pulse contour method and echocardiography and between blood flow/cardiac output estimated using pulse contour method and blood flow/cardiac output estimated using echocardiography (r = 0.84, p < 0.001 and r = 0.87, p < 0.001, respectively). Bland–Altman analysis showed a good agreement (bias −0.20 ± 0.50 L/min) and a low percentage of error (25%) for the cardiac output values estimated by the two methods. The bias between the blood flow/cardiac output ratios obtained with the two methods was 5.19% ± 12.3% (percentage of error = 28.1%). Conclusions: The pulse contour method is a valuable alternative to echocardiography for the assessment of cardiac output and the blood flow/cardiac output ratio in patients treated with veno-venous extracorporeal membrane oxygenation.
2020
Bond, O., Pozzebon, S., Franchi, F., Zama Cavicchi, F., Creteur, J., Vincent, J.-., et al. (2020). Comparison of estimation of cardiac output using an uncalibrated pulse contour method and echocardiography during veno-venous extracorporeal membrane oxygenation. PERFUSION-UK, 35(5), 397-401 [10.1177/0267659119883204].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/1119286
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo