In this contribution a vectorial impedance meter suitable for measurements on two-electrode and three electrode electrochemical sensors is described. The impedance meter is based on basic instruments that are commonly present in an electronics lab (a digital oscilloscope and a signal generator), on a personal computer (PC), and on an 'ad hoc' developed front-end circuit. It can perform impedance measurements up to 5 MHz with a 10% maximum magnitude error and a 10 degrees maximum phase error.

Fort, A., Chiavaioli, F., Lotti, C., Mugnaini, M., Rocchi, S., Vignoli, V. (2009). Laboratory Impedance Meter For Electrochemical Sensors. In AIP Conference (pp.303-305). AIP [10.1063/1.3156532].

Laboratory Impedance Meter For Electrochemical Sensors

Fort, Ada;Mugnaini, Marco;Rocchi, Santina;Vignoli, Valerio
2009-01-01

Abstract

In this contribution a vectorial impedance meter suitable for measurements on two-electrode and three electrode electrochemical sensors is described. The impedance meter is based on basic instruments that are commonly present in an electronics lab (a digital oscilloscope and a signal generator), on a personal computer (PC), and on an 'ad hoc' developed front-end circuit. It can perform impedance measurements up to 5 MHz with a 10% maximum magnitude error and a 10 degrees maximum phase error.
2009
9780735406742
Fort, A., Chiavaioli, F., Lotti, C., Mugnaini, M., Rocchi, S., Vignoli, V. (2009). Laboratory Impedance Meter For Electrochemical Sensors. In AIP Conference (pp.303-305). AIP [10.1063/1.3156532].
File in questo prodotto:
File Dimensione Formato  
C57.pdf

non disponibili

Tipologia: Post-print
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 509.35 kB
Formato Adobe PDF
509.35 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/4593
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo