In this contribution a vectorial impedance meter suitable for measurements on two-electrode and three electrode electrochemical sensors is described. The impedance meter is based on basic instruments that are commonly present in an electronics lab (a digital oscilloscope and a signal generator), on a personal computer (PC), and on an 'ad hoc' developed front-end circuit. It can perform impedance measurements up to 5 MHz with a 10% maximum magnitude error and a 10 degrees maximum phase error.
Fort, A., Chiavaioli, F., Lotti, C., Mugnaini, M., Rocchi, S., Vignoli, V. (2009). Laboratory Impedance Meter For Electrochemical Sensors. In AIP Conference (pp.303-305). AIP [10.1063/1.3156532].
Laboratory Impedance Meter For Electrochemical Sensors
Fort, Ada;Mugnaini, Marco;Rocchi, Santina;Vignoli, Valerio
2009-01-01
Abstract
In this contribution a vectorial impedance meter suitable for measurements on two-electrode and three electrode electrochemical sensors is described. The impedance meter is based on basic instruments that are commonly present in an electronics lab (a digital oscilloscope and a signal generator), on a personal computer (PC), and on an 'ad hoc' developed front-end circuit. It can perform impedance measurements up to 5 MHz with a 10% maximum magnitude error and a 10 degrees maximum phase error.| File | Dimensione | Formato | |
|---|---|---|---|
|
C57.pdf
non disponibili
Tipologia:
Post-print
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
509.35 kB
Formato
Adobe PDF
|
509.35 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11365/4593
Attenzione
Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo