RFId is quickly becoming a pervasive technology inside hospitals. EMI on electrical medical equipment is a concern. Patient safety has to be assured by assessing the electromagnetic compatibility (EMC) between Radio Frequency Identification (RFId) equipment and electrical medical devices. This study tests the effects of an active RFId system, used to track patients, on a critical care electrical medical equipment. 16 devices in five different categories have been tested in a children’s Intensive Care Unit (ICU) and no performance modifications were observed.
Iadanza, E., Dori, F., Miniati, R., Corrado, E. (2010). Electromagnetic Interferences (EMI) from active RFId on critical care equipment. In IFMBE Proceedings (pp.991-994). Berlin : Springer [10.1007/978-3-642-13039-7_251].
Electromagnetic Interferences (EMI) from active RFId on critical care equipment
IADANZA, ERNESTO
;
2010-01-01
Abstract
RFId is quickly becoming a pervasive technology inside hospitals. EMI on electrical medical equipment is a concern. Patient safety has to be assured by assessing the electromagnetic compatibility (EMC) between Radio Frequency Identification (RFId) equipment and electrical medical devices. This study tests the effects of an active RFId system, used to track patients, on a critical care electrical medical equipment. 16 devices in five different categories have been tested in a children’s Intensive Care Unit (ICU) and no performance modifications were observed.File | Dimensione | Formato | |
---|---|---|---|
EMI Full article.pdf
non disponibili
Tipologia:
PDF editoriale
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
1.08 MB
Formato
Adobe PDF
|
1.08 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/11365/1215419