Purpose: In the hard x-ray region, the cross sections for the phase shift of low-Z elements are about 1000 times larger than the absorption ones. As a consequence, phase contrast is detectable even when absorption contrast is minimal or absent. Therefore, phase-contrast imaging could become a valid alternative to absorption contrast without delivering high dose to tissue/human body parts. Methods: To enhance the quality of phase-contrast images without increasing the dose, a possible approach could be the partial deconvolution of the finite source size effects by experimental phase-contrast images. The deconvolution procedure, the authors propose, employs the acquisition of two images on a suitable well-known test sample, one in contact and the other in phase-contrast conditions. Both acquired images are used along with a simulated phase-contrast image (obtained from the test sample in ideal conditions of pointlike source illumination) to correctly retrieve the experimental source distribution function. This information allows a generic experimental phase-contrast image, acquired in the same conditions, to be partially deconvolved by finite source size effects. Results: The performed experimental tests indicate that deconvolved images are equivalent to those which would be obtained with a source 40% smaller than the actual size. In turn, this finding is equivalent to an increase of the "effective" lateral spatial coherence length. The corresponding quality improvement of the phase-contrast imaging is directly deducible by the presence of many Fresnel fringes, much better visible with respect to the original experimental phase-contrast images. Conclusions: The use of a test standard sample, always possible in every experimental setup, to partially deconvolve the finite-size-source blurring effects shows that higher quality phase-contrast images could be readily available, making easier diagnoses and tissue/sample analyses. The method could give, in the future, the possibility to further lower the delivered dose to patients, organs, and tissues when compact room-sized and brilliant microfocus x-ray sources will be available for clinical applications in hospitals. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3560889]

Caro, L.D., Scattarella, F., Tangaro, S., Pelliccia, D., Giannini, C., Bottigli, U., et al. (2011). Deconvolution by finite-size-source effects of x-ray phase-contrast images. MEDICAL PHYSICS, 38(4), 1951-1961 [10.1118/1.3560889].

Deconvolution by finite-size-source effects of x-ray phase-contrast images

Bottigli, U.;
2011-01-01

Abstract

Purpose: In the hard x-ray region, the cross sections for the phase shift of low-Z elements are about 1000 times larger than the absorption ones. As a consequence, phase contrast is detectable even when absorption contrast is minimal or absent. Therefore, phase-contrast imaging could become a valid alternative to absorption contrast without delivering high dose to tissue/human body parts. Methods: To enhance the quality of phase-contrast images without increasing the dose, a possible approach could be the partial deconvolution of the finite source size effects by experimental phase-contrast images. The deconvolution procedure, the authors propose, employs the acquisition of two images on a suitable well-known test sample, one in contact and the other in phase-contrast conditions. Both acquired images are used along with a simulated phase-contrast image (obtained from the test sample in ideal conditions of pointlike source illumination) to correctly retrieve the experimental source distribution function. This information allows a generic experimental phase-contrast image, acquired in the same conditions, to be partially deconvolved by finite source size effects. Results: The performed experimental tests indicate that deconvolved images are equivalent to those which would be obtained with a source 40% smaller than the actual size. In turn, this finding is equivalent to an increase of the "effective" lateral spatial coherence length. The corresponding quality improvement of the phase-contrast imaging is directly deducible by the presence of many Fresnel fringes, much better visible with respect to the original experimental phase-contrast images. Conclusions: The use of a test standard sample, always possible in every experimental setup, to partially deconvolve the finite-size-source blurring effects shows that higher quality phase-contrast images could be readily available, making easier diagnoses and tissue/sample analyses. The method could give, in the future, the possibility to further lower the delivered dose to patients, organs, and tissues when compact room-sized and brilliant microfocus x-ray sources will be available for clinical applications in hospitals. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3560889]
2011
Caro, L.D., Scattarella, F., Tangaro, S., Pelliccia, D., Giannini, C., Bottigli, U., et al. (2011). Deconvolution by finite-size-source effects of x-ray phase-contrast images. MEDICAL PHYSICS, 38(4), 1951-1961 [10.1118/1.3560889].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/49538
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo