The aims of this paper is to measure whether ferromagnetic panels sufficiently reduce the high electromagnetic fields (EMF) to which newborns are exposed in incubators and to which caregivers are exposed when working near the incubators. We measured EMF at mattress level in three neonatal incubators with and without ferromagnetic panels between the electric motor and the mattress. We then measured the EMF at the level of the maximum emission point for caregivers, i.e., near the display panel. The ferromagnetic panels were (a) 5 mm thick iron, (b), (c), (d) respectively, one, two, and three sheets of 0.3 mm thick mu-metal. The weight of iron sheet was 4 g/cm2, and mu-metal 0.2 g/cm2. The use of the ferromagnetic panels significantly reduced the EMF. No significant difference in attenuation was recorded using one, two, or, three sheets of mu-metal, or a single sheet of iron. One, two, and three sheets of mu-metal reduced EMFs by 77%, 82%, and 84.3%, respectively; the reduction with iron was 80%. EMF values measured in incubators were higher than those to which the general population is exposed. The use of ferromagnetic panels significantly reduces the level of EMFs to which neonates and caregivers are exposed.

Bellieni, C.V., Bagnoli, F., Pinto, I., Stacchini, N., Buonocore, G. (2005). Reduction of exposure of newborns and caregivers to very high electromagnetic fields produced by incubators. MEDICAL PHYSICS, 32(1), 149-152 [10.1118/1.1829404].

Reduction of exposure of newborns and caregivers to very high electromagnetic fields produced by incubators

BELLIENI, C. V.;BAGNOLI, F.;BUONOCORE, G.
2005-01-01

Abstract

The aims of this paper is to measure whether ferromagnetic panels sufficiently reduce the high electromagnetic fields (EMF) to which newborns are exposed in incubators and to which caregivers are exposed when working near the incubators. We measured EMF at mattress level in three neonatal incubators with and without ferromagnetic panels between the electric motor and the mattress. We then measured the EMF at the level of the maximum emission point for caregivers, i.e., near the display panel. The ferromagnetic panels were (a) 5 mm thick iron, (b), (c), (d) respectively, one, two, and three sheets of 0.3 mm thick mu-metal. The weight of iron sheet was 4 g/cm2, and mu-metal 0.2 g/cm2. The use of the ferromagnetic panels significantly reduced the EMF. No significant difference in attenuation was recorded using one, two, or, three sheets of mu-metal, or a single sheet of iron. One, two, and three sheets of mu-metal reduced EMFs by 77%, 82%, and 84.3%, respectively; the reduction with iron was 80%. EMF values measured in incubators were higher than those to which the general population is exposed. The use of ferromagnetic panels significantly reduces the level of EMFs to which neonates and caregivers are exposed.
2005
Bellieni, C.V., Bagnoli, F., Pinto, I., Stacchini, N., Buonocore, G. (2005). Reduction of exposure of newborns and caregivers to very high electromagnetic fields produced by incubators. MEDICAL PHYSICS, 32(1), 149-152 [10.1118/1.1829404].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11365/24279
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo